• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Hematoma (1884)
Event Date 01/01/2008
Event Type  Injury  
Manufacturer Narrative
Medtronic cryocath was made aware of this event through a search of literature publications.It was not possible to ascertain specific patient or device information from the literature publication or to match the event with previously reported events.This information is based entirely on journal literature.This event occurred outside the us.All information provided is included in this report.Patient information is limited due to confidentiality concerns.Multiple patients were noted in the article; however, a one to one correlation could not be made with unique product lot numbers.The baseline gender/age characteristic is male/(b)(6) for the patients referenced in the article.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.Referenced article: circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study.J am coll cardiol.2008;52(4):273-8.
 
Event Description
The literature publication reports the following patient complications while using a sheath catheter: there vascular complications consisting of five (5) patients with groin hematomas; all of which were managed "conservatively." the status/location of the sheath catheter is unknown.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6707658
MDR Text Key79889514
Report Number3002648230-2017-00326
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K081049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial
Report Date 07/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/28/2016
Initial Date FDA Received07/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
-
-