ARTHREX, INC. BIO-COMP INTER-FERENCE SCRW; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number AR-5028C-08
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2017
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested/is expected for evaluation but has not yet been received.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainant's event typically caused by not inserting the implant co-axial to the bone tunnel, prying/leveraging the implant, flexing the joint during insertion or improper bone preparation.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.
 
Event Description
It was reported that the screw broke-off during inserting.Surgeon is a high experienced surgeon and is used to use this screw.The rest of the screw was left in the drill channel.There was no harm to the patient, operator or third party.The surgery was finished successfully as half of the screw was left in the drill channel.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP INTER-FERENCE SCRW
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr mdr analyst
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key6714711
MDR Text Key80115043
Report Number1220246-2017-00252
Device Sequence Number1
Product Code MAI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K071176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/01/2018
Device Catalogue NumberAR-5028C-08
Device Lot Number10071747
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-