A healthcare provider reported that a patient had a replacement due to battery depletion.After surgery, the therapy impedance was at 505 ohms.The patient's electrode impedance was 400-500 ohms for all leads.They programmed at 2.5 volts with 2<(>&<)>3 combination being used.When the implant was checked on the day of the report, the day after surgery, the impedance was c<(>&<)>2=2495, c<(>&<)>3=252, 2 <(>&<)>3=2294 ohms.It was reviewed that the issue could be dehydration or a connection issue, but was later confirmed that it was more likely a connection issue since the impedance was much higher than expected.The hcp said they would do imaging and they might do an egd.Palpation was discussed, too.The hcp noted they ran impedance at.8, but it sounded like they ran it at the standard default setting.The patient had nausea and vomiting.No further complications were reported.
|
(b)(4).It was clarified via follow up information that the original report was associated with a separate device.This information was captured in the referenced rr.This device, (b)(4), no longer contains a reportable event.If information is provided in the future, a supplemental report will be issued.
|