• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. CENTRALINK DATA MANAGEMENT SYSTEM; DATA MANAGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. CENTRALINK DATA MANAGEMENT SYSTEM; DATA MANAGER Back to Search Results
Model Number CENTRALINK DATA MANAGEMENT SYSTEM
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2017
Event Type  malfunction  
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site and determined the uninterruptible power supply (ups) had failed.The cse replaced the ups and verified the system is fully operational.The cause of smoke was a malfunction of the ups.This instrument is performing according to specifications and no further evaluation of this device is required.
 
Event Description
The operator of the advia centralink informatics system observed smoke being emitted from the uninterruptible power supply (ups).There were no visible flames or damage to property.Patient samples were not affected and there are no known reports of adverse health consequences due to the smoke being emitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRALINK DATA MANAGEMENT SYSTEM
Type of Device
DATA MANAGER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD
registration number: 8020888
chapel lane, swords co.
dublin, ei
Manufacturer Contact
rolando hualpa
511 benedict ave
tarrytown, NY 10591
9145242963
MDR Report Key6722675
MDR Text Key80339758
Report Number2432235-2017-00425
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCENTRALINK DATA MANAGEMENT SYSTEM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2017
Initial Date FDA Received07/18/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-