MAUDE Adverse Event Report: VYAIRE MEDICAL, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS Back to Search Results
Model Number SIPAP
Device Problems No Audible Alarm (1019); Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 06/13/2017
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical complaint number (b)(4).At this time, vyaire medical has not received the suspect device/component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that when the sipap unit was switched off, there was no audible alarm.The error log showed an e80 (audible alarm module not functioning) error code.It is unknown if there was patient involvement associated with this reported issue.
 
Manufacturer Narrative
The vyaire medical factory service received the suspect device, a sipap ventilator, and evaluated the device.An evaluation of the device duplicated the reported issue and isolated the issue to the main printed circuit board assembly (pcba).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6727283
MDR Text Key80494019
Report Number2021710-2017-06307
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIPAP
Device Catalogue Number27475-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/20/2017
Initial Date FDA Received07/20/2017
Supplement Dates Manufacturer Received08/22/2017
Supplement Dates FDA Received09/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-