|
Model Number SYMPHONY DR 2550 |
Device Problem
Low Battery (2584)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/19/2017 |
Event Type
Injury
|
Event Description
|
Reportedly, the device was last checked on (b)(6) 2017 at which time the battery impedance was 5.67ko and the programmer software provided a time to eri estimate of 17 months.On (b)(6) 2017, the device was found to have reached rrt with a battery impedance of (b)(6).The patient was undergoing treadmill testing during which it was observed the patient¿s cardiac rate was not increasing sufficiently during the exercise so, it was decided to check the pacemaker.When the device was interrogated, it was observed that it was operating in nominal settings due to having reached rrt.It was not suspected that the programming or usage of the device would have changed significantly since the (b)(6) 2017 follow-up.
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Event Description
|
Reportedly, the device was last checked on (b)(6) 2017 at which time the battery impedance was 5.67kohms and the programmer software provided a time to eri estimate of 17 months.On (b)(6) 2017, the device was found to have reached rrt with a battery impedance of 11.42kohms.The patient was undergoing treadmill testing during which it was observed the patient's cardiac rate was not increasing sufficiently during the exercise so, it was decided to check the pacemaker.When the device was interrogated, it was observed that it was operating in nominal settings due to having reached rrt.It was not suspected that the programming or usage of the device would have changed significantly since the (b)(6) 2017 follow-up.
|
|
Search Alerts/Recalls
|
|
|