• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET SPORTS MEDICINE JUGGERKNOT 2.9 W/ TAPERED NEEDLES; FASTENER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET SPORTS MEDICINE JUGGERKNOT 2.9 W/ TAPERED NEEDLES; FASTENER, FIXATION Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problems Unspecified Infection (1930); Nausea (1970); Pain (1994); Swelling (2091); Vomiting (2144); Fluid Discharge (2686); No Code Available (3191)
Event Date 07/18/2017
Event Type  Injury  
Manufacturer Narrative
(b)(6).Product is currently not expected to return to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up medwatch will be submitted.
 
Event Description
It was reported that the patient underwent a shoulder procedure using an anchor and is experiencing sores on hand, nausea, vomiting, pain, swelling and draining at the incision.Patient has also been indicated for revision due to infection.No revision has occurred to date.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent a shoulder repair procedure using an anchor and is experiencing sores on hand, nausea, vomiting, pain, swelling and draining at the incision.The patient is now being indicated for revision due to infection.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent a right shoulder arthroscopy procedure with rotator cuff repair utilizing an anchor, and is experiencing sores on hand, nausea, vomiting, pain, swelling and draining at the incision.The patient is now being indicated for revision due to infection approximately two (2) years post-implantation.
 
Manufacturer Narrative
(b)(4).The reported event could not be confirmed based on limited information received.No products were returned; therefore, the visual and dimensional inspections were not performed.Device history record (dhr) was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.It is noted in initial op-notes that "the juggerknot suture anchor was placed in the bicipital groove.The proximal end of the biceps tendon stump was amputated.It was suspected that the tenodesis site had a good solid repair of the biceps in the anterior humerus.Investigation identified that this product was in scope of a previous investigation that identified the tyvek seal on the product was potentially weak, which compromised sterility and lead to infection and the reported issue.The root cause was determined to be manufacturing deficiency.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT 2.9 W/ TAPERED NEEDLES
Type of Device
FASTENER, FIXATION
Manufacturer (Section D)
BIOMET SPORTS MEDICINE
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6731955
MDR Text Key80669352
Report Number0001825034-2017-05111
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date03/31/2020
Device Model NumberN/A
Device Catalogue Number110005096
Device Lot Number649620
Other Device ID Number(01) 00880304548091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/23/2017
Initial Date FDA Received07/21/2017
Supplement Dates Manufacturer Received07/05/2017
07/24/2016
01/24/2018
05/01/2019
Supplement Dates FDA Received08/02/2017
08/21/2017
01/25/2018
05/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age49 YR
Patient Weight111
-
-