The customer discarded all reagents and so the supposedly defective product was not available for investigational testing.Therefore our quality control laboratory tested the retained samples solidscreen ii controls (sc ii control lot # 80651410-00, scii control b lot # 8637070-00) on tango infinity with the same reagents that were used by the customer.The reagents were tested with biotestcell 3.All positive and negative reactions were correct.We did not observe any false negative reaction.Testing by the quality control laboratory confirmed that the allegedly defective lots of solidscreen ii controls functions correctly.A review of the batch record documentation showed no irregularities which might have negative influences on the quality of the allegedly defective lots.The affected tango infinity was inspected by one of our field service engineers.A partial metrology was performed including performance verification.The check valve for the wash station was noticed to fail and it was replaced.The instrument was confirmed to return operating within specification by the fse.The customer will be informed that the instrument does flag failed qc results and meanwhile, a new flush pump with integrated check valve is available (service bulletin t-223i, distributed on july 18, 2017), which prevents issues with the previous check valve part.
|