BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H802232390012
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Date 06/01/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device evaluated by manufacturer: the device was returned for analysis.Visual inspection of the returned device revealed that the wire is broken 329.3cm from the proximal end, body kinked, spring tip detached and stretched, and a section of the wire was attached to the ball weld of the spring tip.The overall length and outer diameter (od) of the distal tip could not be measured due to device condition (broken wire).The od of the middle and proximal section are within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
 
Event Description
Reportable based on device analysis completed on 03-jul-2017.It was reported that the tip of the wire was bent.During unpacking, the tip of the rotawire¿ and wireclip¿ torquer was noted to be bent.The device was not used and the procedure was competed with another of the same device.There were no patient complications.However, device analysis revealed that the spring tip of the wire was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE¿ AND WIRECLIP¿ TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6740812
MDR Text Key80957248
Report Number2134265-2017-07155
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729195573
UDI-Public(01)08714729195573(17)20190206(10)20254467
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 07/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/06/2019
Device Model NumberH802232390012
Device Catalogue Number23239-001
Device Lot Number20254467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/03/2017
Initial Date FDA Received07/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-