Catalog Number 10-401FC
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Uterine Perforation (2121); Injury (2348)
Event Date 06/27/2017
Event Type  Injury  
Manufacturer Narrative
Lot and serial number of the disposable device not provided by the complainant, therefore the expiration date is not known.The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Lot number of the disposable device not provided by the complainant, therefore the manufacture date is not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.According to the instructions for use (ifu) warnings: to avoid perforation, keep the device tip under direct visualization and exercise care at all times when maneuvering it or cutting it close to the uterine wall.Never use the device tip as a probe or dissecting tool.Reference internal complaint cc#(b)(4).
 
Event Description
It was reported the physician performed a myosure procedure for uterine tissue removal on (b)(6) 2017 and the fluid deficit started to rise.The physician aborted the procedure and performed a laparoscopy which revealed a "large perforation and a superficial wound on the wall of the uterus".The physician stitched the perforation and gel was applied to stop the bleeding.There was no bowel injury noted.The patient was admitted into the hospital overnight for observation and discharged the following day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE REACH HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Type of Device
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
sidra piracha
250 campus drive
marlborough, MA 01752
5082638884
MDR Report Key6747562
MDR Text Key81181257
Report Number1222780-2017-00191
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Catalogue Number10-401FC
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/27/2017
Initial Date FDA Received07/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MYOSURE CONTROLLER: SERIAL NUMBER UNKONWN
Patient Outcome(s) Hospitalization; Required Intervention;
-
-