• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Unable to Obtain Readings (1516); Device Operates Differently Than Expected (2913); Device Contamination with Chemical or Other Material (2944)
Patient Problems Therapeutic Effects, Unexpected (2099); Therapeutic Response, Decreased (2271); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient with an implanted neurostimulator (ins) for urinary dysfunction/sacral nerve stimulation and gastrointestinal /pelvic floor therapy.It was reported the patient¿s ins and lead were removed as therapy was unsatisfactory in resolving symptoms; the system was not working as expected for at least a year.The hcp noticed that upon removal, there was a white milky type substance grown around the battery.Patient services reviewed with caller that this is not caused by a leaking battery and but is biologic material that is a body response to the ins.There were no trauma/falls reported that could be related to this issue and no unrelated medical issues.The hcp checked impedances and everything was normal except contact c/0 were showing ????.No further complications were reported or are anticipated.
 
Manufacturer Narrative
The other applicable components are: product id: 3093-28, lot# v323599, implanted: (b)(6) 2009, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a hcp.It was reported that the cause of the system not working was battery corrosion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6747804
MDR Text Key81186442
Report Number3004209178-2017-15718
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2017
Initial Date FDA Received07/27/2017
Supplement Dates Manufacturer Received08/10/2017
Supplement Dates FDA Received08/23/2017
Date Device Manufactured09/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-