• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P5
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Melted (1385); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has not been returned to olympus medical systems corp.For evaluation.If additional and significant information becomes available, this report will be supplemented.
 
Event Description
Olympus (b)(4) investigated the subject device and it was found that the distal cover was burned / melted and partially missing.It was also found that there was image irregularity with the subject device.There was no patient injury associated with this event reported.
 
Manufacturer Narrative
This supplemental report is being submitted as additional information has been obtained.The manufacturing record of the subject device was reviewed without irregularity related to this event.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 
426425177
MDR Report Key6756219
MDR Text Key82126725
Report Number8010047-2017-01120
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P5
Other Device ID Number04042761066693
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2017
Initial Date FDA Received08/01/2017
Supplement Dates Manufacturer Received08/04/2017
02/13/2019
Supplement Dates FDA Received08/28/2017
02/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-