• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Information (3190)
Event Date 07/20/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A healthcare professional reported that the system froze without system message being displayed before surgery.
 
Manufacturer Narrative
The system was examined and the reported event was replicated.The host module was replaced to resolve the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on october 5, 2011.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming host module.However, how or when the module became nonconforming cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6760790
MDR Text Key81801739
Report Number2028159-2017-03368
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number4.00.168
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/21/2017
Initial Date FDA Received08/02/2017
Supplement Dates Manufacturer Received10/16/2017
Supplement Dates FDA Received10/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-