• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS; OQY INTRAUTERINE BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS; OQY INTRAUTERINE BALLOON Back to Search Results
Catalog Number J-SOSR-100500
Device Problem Partial Blockage (1065)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event Description
The customer reported while preparing the cook bakri postpartum balloon with rapid installation components for bakri placement for hemorrhage, the white part used to spike the bag seemed to be occluded.This was discovered before the bakri was placed.The tubing was unable to be primed, therefore the balloons were not inflated.The bakri was not used on the patient and the procedure was completed using a different bakri.There was no patient harm or consequence.
 
Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, documentation, and specifications of the device were conducted during the investigation.No complaint device has been returned for investigation, and so a physical examination could not be performed.No photos have been provided.Additionally, a document based investigation evaluation was performed.There was no evidence to suggest the product was not made to specifications.A review of the device history record revealed three nonconformances for lot 7674795 were identified and the devices were scrapped prior to further processing.It should be noted there were two other reported complaints associated with this lot number and that all three complaints have the same failure mode from the same customer.Based on the information provided, the actual root cause is unknown and so no conclusion can be drawn.However, appropriate measures have been initiated to address this failure mode.The quality engineering risk assessment for this failure mode was reviewed and it was determined that no additional risk mitigating activity is required at this time.The appropriate personnel will be notified and we will continue to monitor for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS
Type of Device
OQY INTRAUTERINE BALLOON
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6763871
MDR Text Key81732274
Report Number1820334-2017-02160
Device Sequence Number1
Product Code OQY
UDI-Device Identifier10827002242378
UDI-Public(01)10827002242378(17)200214(10)7674795
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberJ-SOSR-100500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2017
Initial Date FDA Received08/03/2017
Supplement Dates Manufacturer Received11/15/2017
Supplement Dates FDA Received12/04/2017
Date Device Manufactured02/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-