The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction of the tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review of kit lot f221 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot f221 for the reported complaint issue shows no trends.Trends were reviewed for complaint categories, alarm #56: red cell pump (#2) error and tubing leak.No trends were detected for these complaint categories.A photo analysis was conducted for this complaint.A review of the photos confirmed the leak.An evaluation of the photos found evidence of a leak within the yellow striped tubing when the tubing is squeezed.The review of the photos indicated that there was a defect on the tubing which could have caused the leak.However, it is not clear from the photos what kind of damage was present on the tubing.Thus the root cause for the leak could not be determined based on the available information.The device history record (dhr) review did not result in any related nonconformances and this kit lot had passed all lot release testing.A material trace of the yellow striped tubing used to build this kit lot also did not find any related nonconformances.All kits under go an occlusions/leak test before they are released.This test identifies any occluded/ leaking tubing or other components within the kit.Only the kits that have passed this test are released for use.As this kit had passed all lot release testing, the damage seen on the tubing is most likely not manufacturing related.The assessment is based on information available at the time of the investigation.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).
|