• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG-OKLAHOMA CITY TRUCLEAR 5.0 OP HYSTERSCOPE; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG-OKLAHOMA CITY TRUCLEAR 5.0 OP HYSTERSCOPE; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 72202976-
Device Problem Difficult to Insert (1316)
Patient Problem No Patient Involvement (2645)
Event Date 07/03/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during testing.The biopsy channel will not allow certain instruments through.There was no patient involvement.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection showed the cover glass is scratched and leaky, distal tip damage, no permeability of working channel, damaged stopcock handle, and particulates under proximal cover glass.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed condition was determined to be a result of a misuse of the product.The ifu recommends " careful inspection of the operative headpiece before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative handpiece.¿ damage may occur if the headpiece is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR 5.0 OP HYSTERSCOPE
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG-OKLAHOMA CITY
75 s. meridian ave
oklahoma OK 73107
Manufacturer (Section G)
MITG-OKLAHOMA CITY
75 s. meridian ave
oklahoma OK 73107
Manufacturer Contact
sharon murphy
60 middletown avenue
north haven, CT 06473
2034925267
MDR Report Key6766658
MDR Text Key81803022
Report Number1643264-2017-21069
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202976-
Device Catalogue Number72202976-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/05/2017
Initial Date FDA Received08/04/2017
Supplement Dates Manufacturer Received01/04/2018
Supplement Dates FDA Received02/02/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-