Device Problems Device Stops Intermittently (1599); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 07/10/2017
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported the unit intermittently indicates system errors after boot-up.There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The carrier/com express board was replaced to resolve the issue.
 
Manufacturer Narrative
The customer troubleshot this issue with ge healthcare technical support.The customer refused onsite service from ge healthcare.There is no further information available on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 
MDR Report Key6773533
MDR Text Key82663606
Report Number2112667-2017-01477
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 09/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2017
Initial Date FDA Received08/08/2017
Supplement Dates Manufacturer Received07/10/2017
07/10/2017
Supplement Dates FDA Received09/07/2017
09/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-