MAUDE Adverse Event Report: OPTONOL LTD. EX-PRESS GLAUCOMA FILTRATION DEVICE; IMPLANT, EYE VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OPTONOL LTD. EX-PRESS GLAUCOMA FILTRATION DEVICE; IMPLANT, EYE VALVE Back to Search Results
Model Number P-50 PL
Device Problem No Flow (2991)
Patient Problem No Code Available (3191)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21cfr 803.56 when additional reportable information becomes available.Results from the product history record review indicated the product met release criteria.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A doctor reported that no flow of aqueous humor was confirmed after shunt implantation and thus it was explanted and a backup product was implanted within the same session.The surgery was then completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EX-PRESS GLAUCOMA FILTRATION DEVICE
Type of Device
IMPLANT, EYE VALVE
Manufacturer (Section D)
OPTONOL LTD.
kiryat hatikshoret
communication center
neve-ilan 90850
IS  90850
Manufacturer (Section G)
OPTONOL LTD.
kiryat hatikshoret
communication center
neve-ilan 90850
IS   90850
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6774666
MDR Text Key82718069
Report Number3003701944-2017-00135
Device Sequence Number1
Product Code KYF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K012852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Model NumberP-50 PL
Device Catalogue Number60053
Device Lot Number072858
Other Device ID Number00380650705013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/11/2017
Initial Date FDA Received08/08/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-