• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 11MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #5 11MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531G511
Device Problems Material Deformation (2976); Scratched Material (3020); Insufficient Information (3190)
Patient Problems Pain (1994); Injury (2348)
Event Date 07/15/2010
Event Type  Injury  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Total knee revision was performed.Date of implant was identified as (b)(6) 2010.
 
Manufacturer Narrative
An event regarding pain involving a triathlon insert was reported.The event was not confirmed.Device evaluation and results: the product was returned in used condition.The images of returned product shows damage on distal end of articulating surface and scraches on many places.Examination of the returned product with material analysis engineer indicated "common damage modes of in-service uhmwpe observed on insert.Explantation damage observed." medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: it was reported that patient had revision surgery due to pain.The event could not be confirmed nor could the root cause be determined because insufficient information was provided.The material analysis of returned device indicated that common damage modes of in-service uhmwpe and damage on distal end of articulating surface are observed on insert.Explantation damage was also observed on the product.A capa trend analysis was conducted for the reported failure mode and concluded pain may result from other factors not necessarily related to the device.No further investigation is possible at this time.Further information such as patient history, histopathology report & follow-up notes are needed to investigate this event further.If additional information and/or device becomes available, this investigation will be reopened.
 
Event Description
Total knee revision was performed.Date of implant was identified as (b)(6) 2010.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #5 11MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
brian lauro
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6779160
MDR Text Key82332466
Report Number0002249697-2017-02432
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/01/2013
Device Catalogue Number5531G511
Device Lot NumberLAZ781
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/13/2017
Initial Date FDA Received08/09/2017
Supplement Dates Manufacturer Received09/26/2017
Supplement Dates FDA Received10/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age64 YR
-
-