• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Over-Sensing (1438)
Patient Problems Twiddlers Syndrome (2114); Impaired Healing (2378)
Event Date 05/01/2017
Event Type  Injury  
Event Description
It was reported that the patient's incision site is not healing a month after implant due to suspected twiddler's syndome.Additionally, the device recorded noise.The device was explanted.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6780284
MDR Text Key82346023
Report Number9614453-2017-02752
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169732216
UDI-Public00643169732216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/14/2018
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/03/2017
Initial Date FDA Received08/09/2017
Supplement Dates Manufacturer Received08/14/2017
Supplement Dates FDA Received10/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age42 YR
-
-