• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A22040A
Device Problems Break (1069); Detachment Of Device Component (1104); Material Fragmentation (1261)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a transurethral resection of the prostate (turp) procedure, the ceramic insulation at the distal end of the inner sheath broke off and fell inside the patient's bladder upon insertion.However, no fragments remained inside the patient since the ceramic insulation was crushed into little pieces which were then flushed out.No further information was provided but there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that the ceramic insulation at the distal end of the resection sheath is broken off completely.Furthermore, the inner sheath is in poor condition with peeling off laser marking, corrosion, contamination and residues due to improper reprocessing.The cause of this damage and the breakage of the ceramic insulation is mechanical overload by the application of excessive force like impact, fall, shock or similar stress.In addition, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the inner sheath without showing any abnormalities.It is clearly stated as a warning note in the instructions that impact, fall, shock or similar stress can damage the ceramic insulation at the sheath's distal end and that the instrument must not be used if damaged since otherwise this can cause injuries to the patient and/or user.The user apparently did not follow these instructions since the damage and breakage of the ceramic insulation were caused by mechanical overload.Therefore, this event/incident was attributed to use error and the case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes.In addition, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, 22045
GM   22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg, 22045
GM   22045
40 66966
MDR Report Key6787391
MDR Text Key83114066
Report Number9610773-2017-10005
Device Sequence Number1
Product Code FJL
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Catalogue NumberA22040A
Device Lot Number157W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2017
Initial Date FDA Received08/11/2017
Supplement Dates Manufacturer Received09/05/2017
Supplement Dates FDA Received09/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-