Catalog Number 20-1037 |
Device Problem
Incorrect Device Or Component Shipped (2962)
|
Patient Problem
No Information (3190)
|
Event Date 07/31/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Upon completion of the investigation, a follow up report will be filed.
|
|
Event Description
|
As reported by the ous affiliate, before the operation started, scrub nurse found 11 pieces in a raney scalp clip 10-pack.No delays were reported; to resolve issue, another pack was opened.
|
|
Manufacturer Narrative
|
Upon completion of the investigation it was noted that no product from this lot is in inventory (produced & shipped september, 2014).True root cause is unknown due to the manufacturing time of the product.Several piece count verifications were within the process with several added over the past few years.Additional scale/piece checks were added to the job instructions.Increased lot inspection of the bagged clips.Continued awareness to team members for heightened inspection of the product along with all the above.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|
|