Model Number 840
Device Problems Occlusion Within Device (1423); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an 840 ventilator generated a safety valve occluded error code and a ventilator inoperable error code.The ventilator was not in use on a patient at the time of the event.The service engineer found that the safety valve needed to be replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6798927
MDR Text Key82877016
Report Number8020893-2017-07474
Device Sequence Number1
Product Code CBK
UDI-Device Identifier10884521077133
UDI-Public10884521077133
Combination Product (y/n)N
Reporter Country CodeBG
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number4-840220DIUA-EN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/19/2017
Initial Date FDA Received08/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-