Varian medical systems informed forte automation systems on july 19, 2017 of a finding at a customer site during commissioning of the proton system.They noted sporadic position inaccuracy of the robotic treatment table during tests at the psi gantry 3 on (b)(6) 2017.The treatment table was moved to a 95-deg yaw position.The table motion was completed with the proton therapy controller [ptc] displaying all coordinates for the intended position.However, the position of a laser tracker target was found to be 14 millimeters off the intended position (isocenter).After moving the table to a different position and going back to exactly the same 95-deg position, the laser tracker target was found at isocenter within the expected tolerances.Varian investigation of the issue was unable to identify a problem with the varian proton therapy controller and questioned whether the problem was with the forte automation patient positioning system.
|