The covering detachment is confirmed.The stent has been crimped down, even though the report states that this happened prior to crimping.The od of the stet is 0.150".The stents are shipped with an od of 0.181".Sever overlapping of the zigs at the ends of the stent were observed (see attached photos).One of the zigs has been manipulated in a way that left it completely turned over.Upon examination of the stent it has been determined that the detachment was most likely caused by mishandling of the stent either during crimping or before crimping.The od of the returned stent was well below the od of stents when they are shipped.The ends of the stent were overlapped and deformed.All covered stents are inspected in final qc to insure each covering has (4) acceptable glue spots.The stent are visually inspected once more after they are packaged for covering adhesion.A sample from each lot is tested for covering strength.The sample from this lot failed at 2.82lbs, well above the 1.5lb minimum.
|