• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NGE-022115
Device Problems Device Inoperable (1663); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event Description
It was reported via the area representative that two (2) ngage nitinol stone extractors did not function at all when they were removed from the package.As reported, the patient was not affected.There were no adverse effects to the patient due to this device issue.Additional patient and event details have been requested but not received at the time of this report filing.
 
Manufacturer Narrative
Additional information: initial reporter name and occupation added.Type of procedure performed: ureteroscopy.The device did not come patient's contact as it was tested prior to actual use.(b)(4).Investigation ¿ evaluation: the ngage nitinol stone extractor was not returned for evaluation and no photographs were provided.Without the complaint device, a physical investigation was not able to be completed.There is no indication that a design or process related failure mode contributed to this event.A review of production and quality documentation did not identify any specific issues with current manufacturing or quality controls that may have contributed to this incident.A review of the device history record could not be completed as the product lot number was not provided.A review of complaint history for the product and lot number combination could not be completed without the lot number.Based on the provided information and the investigation evaluation a definitive root cause could not be determined.Measures have been initiated to address this failure mode.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6804626
MDR Text Key83214377
Report Number1820334-2017-02683
Device Sequence Number1
Product Code FFL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberNGE-022115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/10/2017
Initial Date FDA Received08/18/2017
Supplement Dates Manufacturer Received11/07/2017
Supplement Dates FDA Received11/27/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-