Product evaluation: failure analysis for fiber (b)(4): the fiber/glass cap is partially fractured distal to the uv glue zone; the glass cap distal part is detached and not returned; the heat shrink tubing open end exhibits partially erased red octagonal sign and blue arrow indicator.Based on the device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
|
It was reported that at 12 minutes and 63,358 joules of use while using the side-firing surgical fiber during a prostate procedure, the fiber output beam was observed to be forward-firing.The fiber tip (cap) was cleaned during the procedure.The fiber was replaced and the procedure completed using a second surgical fiber for 99,414 joules.There was no patient injury reported.
|