• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HL CORP DRIVE MEDICAL; MECHANICAL WALKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HL CORP DRIVE MEDICAL; MECHANICAL WALKER Back to Search Results
Model Number 10215RD-1
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Fall (1848); Bone Fracture(s) (1870)
Event Date 06/24/2017
Event Type  Injury  
Event Description
Drive devilbiss healthcare received notification of an incident regarding a rollator that drive imports and distributes.The enduser was attempting to sit down on the rollator when allegedly the right side brake did not enage causing her to fall and break her shoulder.She was taken to mercy hospital carthage for treatment.Drive has evaluated the unit and confirmed the brakes were not engaging when pushed downwards.However, the manual states the brakes can be adjusted to engage properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE MEDICAL
Type of Device
MECHANICAL WALKER
Manufacturer (Section D)
HL CORP
the 3rd industrial park
bitou village, song gang town
baoan district, 51810 5
CH  518105
MDR Report Key6813553
MDR Text Key83461291
Report Number2438477-2017-00069
Device Sequence Number1
Product Code ITJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2017,07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number10215RD-1
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/23/2017
Distributor Facility Aware Date07/26/2017
Event Location Home
Date Report to Manufacturer08/23/2017
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/23/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
Patient Weight181
-
-