• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE CLINITRON RITEHITE C-8 ASSY; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CLINITRON RITEHITE C-8 ASSY; BED, AIR FLUIDIZED Back to Search Results
Model Number P0800010019
Device Problem Insufficient Information (3190)
Patient Problem Skin Erosion (2075)
Event Date 08/10/2017
Event Type  Injury  
Manufacturer Narrative
The hill-rom technician found the proportional valve needed to be replaced.The technician replaced the proportional valve to resolve the issue.The patient was evaluated by the hospital nursing staff and found to have a sacral wound that had progressed from a stage 2 to a stage 3, as well as 3 wounds on her left foot toes at a stage 2 and 1 on the fifth toe at a stage 1.The patient was treated with wet to dry dressings, santyl ointment, hydrogel and skin prep.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.Per the hill-rom user manual, poor fluidization; if fluidization is sluggish or uneven, notify your hill-rom representative.Fluidization is affected by the following: room temperature, humidity, the amount of materials, such as fluid, cells or cellular debris, which has escaped from the blood vessels and has been deposited in tissues or on tissue surfaces, restricted air circulation from blankets on the bed.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2017.It is unknown if the facility performed any other preventative maintenance on this bed.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the patient's sacral wound had worsened from a stage 2 to stage 3.The bed was located at the account.There was a patient injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON RITEHITE C-8 ASSY
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key6820221
MDR Text Key83637912
Report Number1824206-2017-00352
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other Caregivers
Device Model NumberP0800010019
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2017
Initial Date FDA Received08/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
Patient Weight52
-
-