LIVANOVA DEUTSCHLAND S5 CONTROL PANEL MRP 150/85; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 28-95-80
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/31/2017
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 control panel mrp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to reproduce and confirm the reported issue.The issue could be traced to a faulty main switch, which showed incomplete contact.The main switch was replaced and a test-run was successfully performed.The device was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.A supplier quality notification has been opened for a this type of issue.Evaluated on site by livanova rep.
 
Event Description
Livanova (b)(4) received a report that the s5 control panel mast roller pump (mrp) did not function properly during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 CONTROL PANEL MRP 150/85
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6822224
MDR Text Key83747260
Report Number9611109-2017-00665
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number28-95-80
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/31/2017
Initial Date FDA Received08/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-