• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ORG-9100A; MULTIPLE PATIENT RECEIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ORG-9100A; MULTIPLE PATIENT RECEIVER Back to Search Results
Model Number ORG-9100A
Device Problems Radio Signal Problem (1511); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/28/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reported the org experienced signal loss.A transmitter was experiencing signal loss, and testing the device on another org confirmed the issue was with this org.The biomedical engineer was provided with the part number for a replacement org to resolve the issue.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The biomedical engineer reported the org experienced signal loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9100A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6822812
MDR Text Key83731138
Report Number8030229-2017-00273
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103883
UDI-Public04931921103883
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9100A
Device Catalogue NumberORG-9100A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/25/2017
Distributor Facility Aware Date07/28/2017
Device Age105 MO
Event Location Hospital
Date Report to Manufacturer08/25/2017
Initial Date Manufacturer Received 08/25/2017
Initial Date FDA Received08/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-