Catalog Number 03L79-31
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2017
Event Type  malfunction  
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, review of instrument logs and a review of labeling.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Event Description
The customer observed a falsely elevated calcium result on the architect c16000 analyzer.The following data was provided: sid (b)(6) initial 4.13, repeat 3.86, 2.41, 2.42 mmol/l.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY CALCIUM
Type of Device
CALCIUM
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6825559
MDR Text Key83880487
Report Number1628664-2017-00324
Device Sequence Number1
Product Code CJY
UDI-Device Identifier00380740005948
UDI-Public00380740005948
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K981578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2018
Device Catalogue Number03L79-31
Device Lot Number36256UN16
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2017
Initial Date FDA Received08/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-