The patient was undergoing a thrombectomy procedure in the iliac vein using indigo system aspiration catheters 8 (cat8s).During the procedure, the physician connected a cat8 to a penumbra system aspiration pump max 220v (pump max) and turned on the pump max.However, upon turning the pump max on, the physician noticed that air seemed to come through the proximal end of the cat8; therefore, the cat8 was disconnected from the pump max.Upon disconnecting the cat8, the physician noticed a hairline crack where the aspiration tubing is connected to the cat8.Therefore, the cat8 was removed from the patient and the procedure was completed using a second cat8.There was no report of an adverse effect to the patient.
|
The indigo system aspiration catheter 8 (cat8) hub was cracked.The device was kinked and bent along its length.Evaluation of the returned device cat8 revealed the hub was cracked.The hub leaked when attempting to flush saline through the catheter during functional testing.This damage is likely a result of mishandling or overtightening during use.Further evaluation revealed a kink and bends along the length of the catheter.These damages were likely incidental and occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|