• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PFC STEM TRIAL EXTRACT; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PFC STEM TRIAL EXTRACT; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 865226
Device Problem Bent (1059)
Patient Problem No Information (3190)
Event Date 08/02/2017
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
It was reported that the tip was bent.Discovered when set returned to the office for inspection.
 
Manufacturer Narrative
Examination of the returned device confirmed the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC STEM TRIAL EXTRACT
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6836536
MDR Text Key84344375
Report Number1818910-2017-23896
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295216766
UDI-Public10603295216766
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number865226
Device Lot NumberG1010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/02/2017
Initial Date FDA Received08/31/2017
Supplement Dates Manufacturer Received09/25/2017
Supplement Dates FDA Received09/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-