Model Number M00510880 |
Device Problem
Separation Failure (2547)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/07/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
The complainant was unable to report the lot number; therefore, the manufacture date and the expiration date are unknown.However, the complainant reported that the device was not expired.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a trapezoid¿ rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, an alliance handle was used in conjunction with the trapezoid ¿ rx basket in an attempt to crush a stone.However, the basket failed to crush the stone and the basket tip failed to detach.The handle of the trapezoid was cut off and the device was left inside the patient.On (b)(6) 2017, another physician was asked to complete the procedure and used a sohendra lithotripsy system to successfully disengage the tip.A cre dilation balloon was then inserted into the ampulla to create more space in the distal common bile duct.The extractor pro rx balloon was then used to clear the common bile duct from stones and completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ¿stable".
|
|