• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) TRAPEZOID¿ RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) TRAPEZOID¿ RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510880
Device Problem Separation Failure (2547)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/07/2017
Event Type  Injury  
Manufacturer Narrative
The complainant was unable to report the lot number; therefore, the manufacture date and the expiration date are unknown.However, the complainant reported that the device was not expired.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid¿ rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, an alliance handle was used in conjunction with the trapezoid ¿ rx basket in an attempt to crush a stone.However, the basket failed to crush the stone and the basket tip failed to detach.The handle of the trapezoid was cut off and the device was left inside the patient.On (b)(6) 2017, another physician was asked to complete the procedure and used a sohendra lithotripsy system to successfully disengage the tip.A cre dilation balloon was then inserted into the ampulla to create more space in the distal common bile duct.The extractor pro rx balloon was then used to clear the common bile duct from stones and completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ¿stable".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID¿ RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6837235
MDR Text Key84309201
Report Number3005099803-2017-02562
Device Sequence Number1
Product Code LQC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510880
Device Catalogue Number1088
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2017
Initial Date FDA Received09/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-