• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 TEGRESS URETHRAL IMPLANT; TEGRESS¿ URETHRAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 TEGRESS URETHRAL IMPLANT; TEGRESS¿ URETHRAL IMPLANT Back to Search Results
Catalog Number 653001
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Erosion (1750); Incontinence (1928); Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
The sample was not returned.The finished product met all specifications prior to being released for general distribution.The instructions for use which accompanies all devices currently addresses potential risks associated with surgically implanted materials.The instructions for use states in the adverse events: "the tegress¿ urethral implant clinical trial involved 374 tegress¿ implant treatment injections in 174 subjects (mean follow-up of approximately 14 months).There were no deaths among study patients.The following table lists the treatment related adverse events reported during the clinical study (incidence >2%).Treatment related events are those events that were deemed to be related to either the device or the procedure.All genitourinary events were classified as treatment related.Number (%) subjects reporting treatment related adverse events event category tegress¿ implant (n=174) urinary tract infection (uti) 50 (29%) delayed voiding 32 (18%) dysuria 31 (18%) exposed material 28 (16%) urinary urgency 24 (14%) urinary frequency 22 (13%) genitourinary (infection, tenderness) 20 (11%) hematuria 19 (11%) urge incontinence 16 ( 9%) worsening of incontinence (onset of urge) 14 ( 8%) outlet obstruction 13 ( 7%) pain at injection site 13 ( 7%) pelvic pain 13 ( 7%) yeast infection 12 ( 7%) leakage of urine/stress incontinence 9 ( 5%) bulking material injected into bladder 7 ( 4%) fatigue 3 ( 2%) abnormal urinalysis 3 ( 2%) bladder fullness 3 ( 2%) nocturia 3 ( 2%) pelvic heaviness 3 ( 2%) uterine fibroids 3 ( 2%) other (<2%)* 38 (n/a)* other treatment related adverse events in tegress¿ urethral implant patients, occurring at frequencies of < 2%, were as follows (listed alphabetically): abdominal upset, bladder spasms, bladder stones, body aches, burning pain, cold and shivering, cyst, cystitis, feeling of decreased sensation with urination, feeling of bladder not emptying, felt faint during bulking injection, fever, garlic odor, genital pain, kidney stones, labia with erythema, lower back pain, medicinal smell to urine, nausea, partial urinary retention due to dementia, pyuria, removal of tegress¿ implant, urethral burning sensation, urethral irritation, urethral redness, urethral soreness, urethral spasm, vaginal bleeding, and vulvar burning.Most treatment related adverse events occurred within 24 hours of treatment and subsequently resolved within 30 days.At the time of database closure, 92% of treatment related adverse events were resolved.The following events were persistent or resolution was unconfirmed at the time of database closure (the number of events is shown in parentheses): urge incontinence (6); leakage of urine/stress incontinence (5); worsening of incontinence (onset of urge) (5); exposed material (4); uterine fibroids (3); urinary tract infection (2); urinary frequency (2); urinary urgency (2) and one event each of genitourinary (infection/tenderness), kidney stones, nocturia, pelvic pain, urethral redness, and tegress¿ implant removal.Of the treatment related adverse events, 39% were classified as mild, 58% were classified as moderate, and 3% were classified as severe.The severe treatment related adverse events included: bladder spasms, bladder stones, bulking material injected into the bladder, delayed voiding, exposed bulking material, hematuria, pelvic pain, urge incontinence, and urinary frequency.The clinical study consisted of a pilot or feasibility phase(n=28), followed by the expanded study phase (n=146).There was higher overall rate of genitourinary adverse events in the feasibility phase of the study than in the expanded phase.In particular, the rate of ¿exposed material¿ was higher (32% vs.13%).This reduction in the rate of exposed material was achieved as a result of modifications to the tegress¿ implant injection instructions and investigator training.During the course of the clinical investigation (both study phases), 28 subjects (16%) receiving tegress¿ urethral implant treatment experienced exposed bulking material in the urethral mucosa.Patients experiencing exposed material often reported other events, particularly dysuria, delayed voiding, urinary tract infection, hematuria, urinary frequency, and urinary urgency.Exposed tegress¿ implant material was associated with shallow placement and injection too proximal to the bladder neck.Over time, the urethra healed spontaneously as the mucosal surface re-epithelialized.The physician may choose to remove exposed material cystoscopically with graspers or forceps to facilitate healing.The majority of patients were injected via the transurethral approach, while a small proportion of patients were injected periurethrally.There were significantly more adverse events among tegress¿ implant patients treated periurethrally; as a result, the tegress¿ urethral implant instructions for use are limited to transurethral administration.Potential adverse events: although not reported in the clinical study, other potential adverse events which may occur include erosion, erythema, embolic phenomena, and vascular occlusion." (b)(4).Sample not returned.
 
Event Description
It was reported in the patient¿s medical records that as a result of having the product implanted, the patient has experienced urinary tract infection, recurrent incontinence, urinated a ¿big chunk of tan material¿ that the surgeon felt was tegress material, wears pads, intrinsic sphincter deficient, bulking agent erosion into urethra, autologous pubovaginal sling, martius flap, urethral erosion repair, explant of uretex for synthetic mesh exposure, granulation tissue in the periurethral space where the mesh was implanted, suprapubic tube placement, widening of the bladder neck and urethra and self-catheterizations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEGRESS URETHRAL IMPLANT
Type of Device
TEGRESS¿ URETHRAL IMPLANT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
angela robinson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6838184
MDR Text Key84367364
Report Number1018233-2017-04589
Device Sequence Number1
Product Code LNM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030030 S005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Attorney
Type of Report Initial
Report Date 09/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2009
Device Catalogue Number653001
Device Lot Number43CQH001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Device Age9 MO
Event Location Hospital
Initial Date Manufacturer Received 08/12/2017
Initial Date FDA Received09/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TEGRESS¿ URETHRAL IMPLANT
Patient Outcome(s) Required Intervention;
Patient Age74 YR
-
-