Model Number 3058
Device Problems High impedance (1291); Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with a neurostimulator (ins) for urinary dysfunction/sacral nerve stim.It was reported that during a battery replacement, they were getting high impedance on all electrode pairs with 0.It was noted that they were testing with the new ins, and that testing with the old one, prior to surgery had not been possible due to that battery being completely depleted.Measurements for the other electrode pairs were given to be: c-1, ???; c-2, 668; c-3, 668; 1-2, 739; 1-3, 1012; and 2-3, 1012.Additional information was received on (b)(6) 2017.It was reported that the lead was still implanted and connected to the new ins.It was noted that the patient was programmed around electrode 0.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6839908
MDR Text Key84666320
Report Number3004209178-2017-18656
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2019
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/25/2017
Initial Date FDA Received09/01/2017
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-