BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/11/2017
Event Type  malfunction  
Manufacturer Narrative
The exact age of the patient is unknown.However, it was reported the patient was over 18 years.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold¿ lite was used during an anterior prolapse repair procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the needle detached from the suture during placement.The detached piece remained inside the capio head.The procedure was completed with another uphold¿ lite.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual examination of the returned uphold¿ lite revealed that the suture on the blue dilator was broken.The remainder of the suture with the dart was not returned.The suture on the blue with white stripe dilator is cut most likely to facilitate removal.Analysis also revealed that there was no damage to the capio slim suture capturing device.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
 
Event Description
It was reported to boston scientific corporation that an uphold¿ lite was used during an anterior prolapse repair procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the needle detached from the suture during placement.The detached piece remained inside the capio head.The procedure was completed with another uphold¿ lite.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6841193
MDR Text Key85602687
Report Number3005099803-2017-02616
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public(01)08714729839200(17)20190602(10)0000032829
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/02/2019
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000032829
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/14/2017
Initial Date FDA Received09/05/2017
Supplement Dates Manufacturer Received10/04/2017
Supplement Dates FDA Received10/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-