Model Number AP31801-GB
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Additional information will be provided upon conclusions of the investigation.
 
Event Description
Arjohuntleigh has been informed that a system 2000 bath has been tested (b)(6) for contamination with pseudomonas.The bath has been taken out of use.No patient was involved.
 
Manufacturer Narrative
Arjohuntleigh have acknowledged that pseudomonas aeruginosa is extremely common bacteria that is present in moist environments such as soil, water, sinks and showers and often found in spas or purified water system operators.Pseudomonas grows in water and thrives at warm temperatures.It grows quickest if the water is allowed to sit without movement.The investigation including analysis of potential sources of contamination allowed to exclude the manufacturing process as the source of bacteria.The results of microbiological test performed on swabs taken from cold, hot inlets, tap and filter were negative (presumptive pseudomonas unconfirmed).The test of the water supply use during manufacture's quality control process did not reveal any presence of bacteria.Further investigation is ongoing.A follow-up report will be provided as soon as the conclusion of the investigation would become available.
 
Manufacturer Narrative
On 2017-aug-23 arjohuntleigh has been informed by customer (b)(6) (facility management partner) that based on the results of microbiological tests of arjohuntlegh system 2000 bathtub (serial number (b)(4)) located in (b)(6) hospital ((b)(6)) it is contaminated with excessive levels of pseudomonas aeruginosa bacteria.No adverse event or injury was reported.The review of similar reportable events with the involvement of the system 2000 baths in last 5 years, revealed a low number of cases where it was indicated that the bath was contaminated with the pseudomonas aeruginosa bacteria.The occurrence rate observed for this failure mode is considered to be low.Please note that pseudomonas aeruginosa is a common environmental organism and can be found in faeces, soil, water and sewage.It can multiply in water environments and also on the surface of suitable organic materials in contact with water.It can be spreaded by equipment that gets contaminated and is not properly cleaned and maintained.On 2017-aug-23 an arjohuntleigh representative was informed about the microbiological tests performed in (b)(6) hospital on the system 2000 bathtub, which revealed excessive levels of pseudomonas aeruginosa bacterium.According to this information the samples for tests were taken on (b)(6) 2017 from the bathtub (b)(4) located on the ward (b)(6) hospital and were examined by external laboratory services provider.On 2017-jun-20 the facility was informed that the results of tests were positive with pseudomonas bacteria levels of >100 cfu/100ml pre flush and 8 cfu/100ml post flush.The parts of the bathtub from which the samples were obtained remain unknown.The laboratory carried out chlorination on the shower installation of the bathtub to decrease the contamination.No further test results for this bathtub were provided by customer.On 2017-aug-24 the arjohuntleigh received a request from (b)(6) regarding removal of the shower attachment (handle and hose) from the bathtub (b)(4) as a result of bacterium remaining in this part.The customer removed disinfection handle and hose on its own earlier as per recommendation of the internal facility microbiologist consultant.The shower attachment was removed from the bathtub and device was used without the showering function despite that this solution was not approved by arjohuntleigh.The customer was informed by arjohuntleigh representative that this kind of modification is not authorized and may impact the product performance.Please note that according to recommendations described in the operating and product care instructions (04.Ar.07_7gb) which was delivered with this device, the bath should be disinfected immediately after every use and every day.Moreover, it states that user of the bathtub is obligated to remove and clean strainers in the shower heads on monthly basis.In summary, the complaint was decided to be reportable based on the allegation that arjohuntleigh bath system 2000 was contaminated with bacteria.It should be underlined that this bathtub was manufactured in 2006 therefore the manufacturing process cannot be considered as source of contamination with pseudomonas aeruginosa due to long time in use during which device was exposed to various environmental factors.Moreover, it was confirmed that bacterium was found in other points of the facility, which were not adjacent to the arjohuntleigh bathtub.The results of the investigation performed enable us to determine that allegation reported does not compromise patient safety and is not likely to cause or contribute to a death, serious injury or deterioration in the state of health.Therefore this type of event will no longer be seen as a reportable complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 2000
Type of Device
BATH, HYDRO-MASSAGE
Manufacturer (Section D)
ARJO HOSPITAL EQUIPMENT AB
verkstadsvagen 5
eslov, 24121
SW  24121
Manufacturer (Section G)
ARJO HOSPITAL EQUIPMENT AB
verkstadsvagen 5
eslov, 24121
SW   24121
Manufacturer Contact
kinga stolinska
ks. wawrzyniaka 2
komorniki, 62-05-2
PL   62-052
MDR Report Key6841383
MDR Text Key86576084
Report Number3007420694-2017-00185
Device Sequence Number1
Product Code ILJ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup,Followup
Report Date 11/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberAP31801-GB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/27/2017
Distributor Facility Aware Date08/23/2017
Device Age10 YR
Event Location Hospital
Date Report to Manufacturer11/27/2017
Initial Date Manufacturer Received 08/23/2017
Initial Date FDA Received09/05/2017
Supplement Dates Manufacturer Received10/05/2017
11/27/2017
Supplement Dates FDA Received10/31/2017
11/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/25/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-