On (b)(6) 2017, a gore® excluder® aaa endoprosthesis featuring c3® delivery system was being prepared for advancement when resistance was reportedly noted when loading the device onto the guidewire.It was reported the device was not tracking well on the guidewire and was very difficult to advance.It was reported a decision was made to remove the device from the guidewire, but during retraction the device could not be removed.Force was reportedly applied to remove the device from the guidewire, after which point the catheter began to separate at the trailing olive / polyimide wire junction.It was reported that after the catheter started to separate, it was able to be removed from the guidewire.No obstructions were reportedly noted within the catheter itself.It was reported there were no kinks noted in the guidewire after removal of the delivery catheter.However, it was decided to use a new guidewire, and another device was able to be advanced and was successfully implanted.The procedure was completed with no adverse events to the patient.
|