• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH ARTIS ZEE BIPLANE; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH ARTIS ZEE BIPLANE; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 10094141
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.(b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee biplane system.During a clinical procedure, the user reported the table performed unrequested movements when the flat detector was in the front position.As a result, the patient was safely removed from the system and transferred to an alternate system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.During the investigation it was determined that the reference switch for the table tilt was misaligned and falsely activated resulting in the described longitudinal movement.Following readjustment of the reference switch, the system recovered to normal operation.The manufacturer is not considering further actions resulting from this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE BIPLANE
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6844725
MDR Text Key84995560
Report Number3004977335-2017-92940
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K073290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094141
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/06/2017
Event Location Hospital
Date Report to Manufacturer08/22/2017
Initial Date Manufacturer Received 08/22/2017
Initial Date FDA Received09/06/2017
Supplement Dates Manufacturer Received12/06/2017
Supplement Dates FDA Received12/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-