• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40.2
Device Problems No Display/Image (1183); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/10/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported to merge technical support that the provided test plug was used and the unit still did not connect to the hemo monitor pc.It was further confirmed that the pdm powered up correctly and all the connections were checked.A technician was deployed to the site to replace the faulty unit.Testing was performed and it was confirmed that the replaced unit worked as expected.The faulty device was sent to the manufacturer for evaluation and testing.The unit was received by merge healthcare on 23aug2017.The manufacturer's evaluation results showed that the unit's power connector was faulty and subsequently replaced.(b)(4).Methods code: actual device evaluated.Results code: power source problem.Conclusions code: device difficult to operate.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that the pdm (patient data module) would not connect to the hemo monitor pc prior to a procedure starting.Subsequently, the procedure was completed using a third party monitoring system.The brand and type of equipment was not divulged to merge healthcare.Information obtained found that the site has multiple labs but chose to perform the procedure in this lab anyway.With merge hemo not capturing physiological data, there is a potential for incorrect treatment that could result in harm to the patient.The customer reported that procedure was completed successfully with the external monitoring system.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge drive
hartland, WI 53029
2623670700
MDR Report Key6847509
MDR Text Key85746221
Report Number2183926-2017-00175
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2017
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2017
Initial Date FDA Received09/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-