Catalog Number 865226 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Information (3190)
|
Event Date 08/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that while trying to remove the femoral sleeve trial and trial stem, the trial stem extractor would not thread into the trial stem.We had another one in hand and used that fine.When we examined the first stem extractor, the tip where the threads are was noted to be bent.Apparently bent enough that it wouldn't let the doctor thread it into the trial stem.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Additional narrative: examination of the returned device confirmed the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|