• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number UNAVAILABLE
Device Problem Insufficient Information (3190)
Patient Problem No Code Available (3191)
Event Date 08/04/2017
Event Type  Injury  
Manufacturer Narrative
Patient's date of birth unavailable.Device lot, expiration date and complete udi number unavailable from facility device manufacture date unavailable.
 
Event Description
Patient showed no decline of status immediately following a lead extraction procedure; however, a week later a ''shunt'' was discovered between the right and left atrium when patient underwent echocardiogram for an unrelated issue.According to the report, it was thought to have occurred at the time of extraction of the lv lead from the cs.Patient survived the event.
 
Manufacturer Narrative
Additional information provided by physician to warrant correction of the mdr.Manufacturer received word from physician, who provided additional case detail.From this detail, the glidelight device listed in the initial mdr, was reported by the physician to not have been in the area of the coronary sinus where the lv (left ventricular) lead was present.The physician confirmed that an lld was present within the lead in the coronary sinus at the time the lead dislodged.This follow up corrects the device (originally the glidelight) to the lld as being present in the coronary sinus within the lead.The shunt was discovered a week after the procedure was completed.No model, lot, expiration date or udi available for the lld.The lld device was not returned for evaluation.Device manufacture date of the lld unavailable.
 
Event Description
Additional information received from physician regarding event: procedure commenced to remove two rv leads and one lv lead.Dr.Used glidelight device in the subclavian, svc, and ra.One rv lead successfully removed with laser; second rv lead successfully removed without any device needed.Dr.Was not able to enter the coronary sinus due to significant binding of fibrotic tissue.During removal of the rv leads, the lv lead dislodged.Md confirmed an lld was present in each of the leads, and that traction was being applied to the lv lead when it dislodged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key6852549
MDR Text Key85431960
Report Number1721279-2017-00201
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNAVAILABLE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator No Information
Device Model NumberUNAVAILABLE
Device Catalogue NumberUNAVAILABLE
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2017
Initial Date FDA Received09/08/2017
Supplement Dates Manufacturer Received02/09/2018
Supplement Dates FDA Received02/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SPECTRANETICS CVX-300 EXCIMER LASER; SPECTRANETICS LEAD LOCKING DEVICE; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age40 YR
Patient Weight75
-
-