TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 816571 |
Device Problems
Mechanical Jam (2983); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Per biomed, he checked the upper bearing and saw that there was some gunk.He cleaned the roller pump so it's been a little better.The unit will be sent back to the manufacturer for evaluation.
|
|
Event Description
|
After use of the device for a cardiopulmonary bypass procedure, the user reported that the roller pump gave a squeaking sound when it was turned on and sometimes it jammed.There were no reported adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed belt slip error message and intermittent squeaking followed by pump jam error messages as a result of defective bearing.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|