• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH ARTIS ZEEGO; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH ARTIS ZEEGO; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 10280959
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Patient Involvement (2645)
Event Date 08/22/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the axiom artis zeego system.During an interventional procedure, the robot arm was moved causing it to collide with the omnicell cabinet and crack the glass.The procedure was completed and there is no report of impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause has been determined as improper workmanship and misconfiguration of the system during installation.To keep the probability of a collision to a minimum, a collision computer continuously monitors the positions of all integrated components of the system with respect to one another and the room.Under normal circumstances the collision computer detects when configured distances are violated.In the present case the cse found that after the installation of the omnicell cabinet, the room was not reconfigured accordingly.Therefore the collision computer could not detect the reported collision.After the room configuration was corrected and the patient transfer position of the robot has been re-adjusted; a potential collision with the omnicell cabinet can now be detected by the system.No systematic error was identified in the field.Therefore no field corrective action is planned at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEEGO
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6857360
MDR Text Key85783054
Report Number3004977335-2017-93011
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10280959
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/22/2017
Event Location Hospital
Date Report to Manufacturer08/22/2017
Initial Date Manufacturer Received 08/22/2017
Initial Date FDA Received09/12/2017
Supplement Dates Manufacturer Received01/31/2018
Supplement Dates FDA Received01/31/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-