Model Number N/A |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/14/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a tibial nail case, the guide wire gripper will no longer grip the guide wire.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
(b)(4).The reported event could not be confirmed based on limited information received.No products were returned; therefore, the visual and dimensional inspections were not performed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required.A definitive root cause cannot be determined with the information provided.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.The visual inspection noted impact marks and scratches on the impaction surface.Both the holes exhibit heavy gauging and wear indicative of use.The wear on the holes leads to insufficient gripping of guide wires.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The root cause can be determined as normal wear during the instrument¿s field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|