• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ZNN GUIDE WIRE GRIPPER; PASSER, WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ZNN GUIDE WIRE GRIPPER; PASSER, WIRE Back to Search Results
Model Number N/A
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/14/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a tibial nail case, the guide wire gripper will no longer grip the guide wire.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
(b)(4).The reported event could not be confirmed based on limited information received.No products were returned; therefore, the visual and dimensional inspections were not performed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required.A definitive root cause cannot be determined with the information provided.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.The visual inspection noted impact marks and scratches on the impaction surface.Both the holes exhibit heavy gauging and wear indicative of use.The wear on the holes leads to insufficient gripping of guide wires.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The root cause can be determined as normal wear during the instrument¿s field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZNN GUIDE WIRE GRIPPER
Type of Device
PASSER, WIRE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 west center street
po box 708
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6861248
MDR Text Key86133409
Report Number0001822565-2017-06405
Device Sequence Number1
Product Code HXI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Hospital Service Technician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00249001200
Device Lot Number63017696
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/15/2017
Initial Date FDA Received09/13/2017
Supplement Dates Manufacturer Received02/05/2018
04/18/2018
04/18/2018
Supplement Dates FDA Received02/05/2018
04/18/2018
04/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-