Catalog Number 850000010 |
Device Problems
False Negative Result (1225); Incorrect Or Inadequate Test Results (2456)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This is our initial report on this incident.
|
|
Event Description
|
The customer reported a discrepant result when testing the product ahg igg scii on tango infinity # (b)(4) and tango infinity sn (b)(4).A known positive sample showed a 2+ positive reaction on tango infinity # s/n (b)(4) whereas the same sample showed a negative reaction when tested on sn (b)(4).The customer returned the supposedly defective product ahg igg sc ii and a bottle of mlb2 for investigational testing.The sample that had caused false negative test result was not available.Therefore our quality control laboratory tested the supposedly defective product anti-human-globulin, anti-igg solidscreen ii in comparison with the retained lot of this product.The complained reagent showed functionality as expected.Furthermore dat testing was carried out in parallel on two instruments.Therefore dat positive samples were tested on tango infinity and on tango optimo by using the complained reagent from the customer in parallel to the retained sample.All results were comparable and showed clearly positive reactions.Testing by our quality control laboratory confirmed the correct functionality of the allegedly defective lot of anti-human-globulin, anti-igg solidscreen ii.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The affected tango infinity (b)(4) was inspected by one of our field service engineers by performing metrology qualification procedure as required.At this time a cause for an instrument malfunction could not be identified, the evaluation of further data is still ongoing.
|
|
Manufacturer Narrative
|
This is our final report on this incident.
|
|
Event Description
|
The customer reported a discrepant result when testing the product ahg igg scii on tango infinity #(b)(4) and tango infinity sn #(b)(4).A known posiitve sample showed a 2+ positive reaction on tango inifinity # s/n (b)(4) whereas the same sample showed a negative reaction when tested on sn # (b)(4).The customer returned the supposedly defective product ahg igg sc ii and a bottle of mlb2 for investigational testing.The sample that had caused false negative test result was not available.Therefore our quality control laboratory tested the supposedly defective product anti-human-globulin, anti-igg solidscreen ii in comparision with the retained lot of this product.The complained reagent showed functionality as expected.Furthermore dat testing was carried out in parallel on two instruments.Therefore dat positive samples were tested on tango infinity and on tango optimo by using the complained reagent from the customer in parallel to the retained sample.All results were comparable and showed clearly positive reactions.Testing by our quality control laboratory confirmed the correct functionality of the allegedly defective lot of anti-human-globulin, anti-igg solidscreen ii.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The affected tango infinity # (b)(4) was inspected by one of our field service engineers.All points and checks passed.Also multiple dat testing of patient sample in question was done with similar positive results on both instruments.After performing post service verification procedure as required the instrument was confirmed to operate within specification.The customer stated that the problem did not return images of the reported false negative result were not saved and could not be found by customer.Other delivered images from both instruments are showing positive dat results as expected.The initially reported false negative result was observed one time which gave a weak result by tube testing.The tango infinity is not validated for cord blood samples.No indication for an instrument malfunction could be identified.The instrument is working within specifications.
|
|
Search Alerts/Recalls
|