ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL W/ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Entrapment of Device (1212)
Patient Problem Eye Injury (1845)
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: the product was not returned for analysis.Complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.A root cause has not been identified.(b)(4).
 
Event Description
An ophthalmic surgeon reported that sometimes during intraocular lens (iol) implant procedures, the plunger tip of the preloaded iol delivery system gets caught on the iris as she is withdrawing it from the eye after lens implantation.Additional information has been requested.There are two medical device reports associated with this facility for this issue.This report represents an unknown number of patients and procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL W/ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6867796
MDR Text Key86737809
Report Number1119421-2017-01134
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAU00T0
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2017
Initial Date FDA Received09/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-