A customer in switzerland reported a misidentification of campylobacter coli as neisseria cinerea with the vitek® 2 nh test kit (reference 21346).An investigation was performed.The customer reported setting up the strain from campylosel agar and incubating at 37°c in a microaerophilic environment for 24 hours.Campylosel agar is not a recommended agar for use with the nh card.One lab report was submitted showing a very good identification of n.Cinerea with three (3) atypical negative reactions (dmlt, pvate, lglm) for an identification of c.Coli according to the nh knowledge base.An increased number of atypical negative results can indicate a strain with decreased viability, user set up error or an atypical strain.Use of non recommended media may result in reactions which are not supported by the nh knowledge base.Without the strain or raw data, it's not possible to further evaluate the cause of the misidentification.The vitek 2 nh lot #2450361203 met final qc release criteria and passed qc performance testing.
|